  • Ambulatory syringe pump that runs on either AC current or on rechargeable internal batteries (on-screen voltage test is available)
  • Accepts a broad range of syringe brands and sizes, and has automatic sensing of syringe size
  • Medfusion® 2010 has no bolus function, but it can be programmed for continuous, auto or manual intermittent, or Volume/Time infusions
  • Recommended for intravenous or arterial infusions of blood, lipids, fluids, antibiotic, and any medication requiring precisely-controlled infusion rates, including epidural infusions
  • 4-position variable occlusion settings
  • Custom program mode allows preselection of certain modes and parameters, which can then be locked-out by the programming clinician to limit program access
  • Syringe manufacturer can also be custom-programmed if not using a syringe by BD, Monoject, Abbott (50cc only), or Terumo
  • It is also indicated for use with BD glass syringes (1, 3, 5, or 10cc sizes)
  • Medfusion® 2010 can be pole-mounted, using the pole clamp included with the pump
  • Easy-to-read LCD screen with LED alarm and function lights. Accuracy is ± 3 %, excluding variations in syringe accuracies

Specifications:

  • Flow Rate: 0.1-378.0 ml/hr., depending on syringe size selected
  • Volume Limit: 00.1 to max. capacity of syringe in increments of 0.1 ml
  • Volume Delivered: 000.0000 to 9999.9999 mg or 0000.00 to 9999.99 ml
  • Bolus Rate: Up to 378.0 ml/hr
  • KVO: 0.1 - 9.9 ml/hr in increments of 0.1 ml
  • Occlusion: Normal - max. 16.4 PSI with Mono 35 cc syringe; Low - max. 5.96 PSI with Mono 35 cc syringe; Middle - max. 13.8 PSI with Mono 35 cc syringe; High - max. 15.0 PSI with Mono 35 cc syringe
  • Accuracy: ±3%, excluding variations in syringe accuracies
  • Battery Capacity: 16-hour charge operates the pump for approximately 10 hours at 5.0 ml/hr with a 60 ml syringe
  • Recharge Time: 16 hours maximum
  • Power Requirements: AC Adapter or internal battery
  • Dimensions: 4.5" x 3" x 7.5"
  • Alarms: Near Empty, Empty, Volume Limit, Occlusion, Systems Malfunction, Low Battery, Depleted Battery, Syringe Not Loaded, Check Clutch, Invalid Syringe Size

 